QA PROGRAM AND TESTING | Document Reference | Y | N | NA | Comments |
27. | Sampling/sample protocols must be consistent with chapter 314-55 WAC, written and approved by the laboratory director, and must include documented training. | - | - | - | - | - |
a. | Demonstrate adequacy of the chain-of-custody, including: Tracking upon receipt of sample including all personnel handling the sample and documenting condition of the sample through a macroscopic and foreign matter inspection. | - | - | - | - | - |
b. | Macroscopic and foreign matter inspection - Fit for purpose test. Scientifically valid testing methodology: Either AHP monograph compliant or other third-party validation. | - | - | - | - | - |
c. | Failed inspection of product: Tracking and reporting. | - | - | - | - | - |
d. | Return of failed product documentation and tracking. | - | - | - | - | - |
e. | Disposal of used/unused samples documentation. | - | - | - | - | - |
f. | Sample preparation, extraction and dilution SOP. | - | - | - | - | - |
g. | Demonstration of recovery for samples in various matrices (SOPs): | - | - | - | - | - |
i. | Plant material - Flower; | - | - | - | - | - |
ii. | Edibles (solid and liquid meant to be consumed orally); | - | - | - | - | - |
iii. | Topical; | - | - | - | - | - |
iv. | Concentrates. | - | - | - | - | - |
28. | Data protocols. | - | - | - | - | - |
a. | Calculations for quantification of cannabinoid content in various matrices - SOPs. | - | - | - | - | - |
b. | Determination of the range for reporting the quantity (LOD/LOQ) data review or generation. | - | - | - | - | - |
c. | Reporting of data: Certificates of analysis (CA) - Clear and standardized format for consumer reporting. | - | - | - | - | - |
d. | Each test report shall include at least the following information, unless the laboratory has valid reasons for not doing so: | - | - | - | - | - |
i. | A title (e.g., "Test Report" or "Certificate of Analysis"); | - | - | - | - | - |
ii. | The name and address of the laboratory, and the location where the tests were carried out, if different from the address of the laboratory; | - | - | - | - | - |
iii. | Unique identification of the test report certificate (such as the serial number), and on each page an identification in order to ensure that the page is recognized as a part of the test report or calibration certificate, and a clear identification of the end of the test report or calibration certificate; | - | - | - | - | - |
iv. | The name and address of the customer; | - | - | - | - | - |
v. | Identification of the method used; | - | - | - | - | - |
vi. | A description of, the condition of, and unambiguous identification of the item(s) tested; | - | - | - | - | - |
vii. | The date of receipt of the test item(s) where this is critical to the validity and application of the results, and the date(s) of performance of the test or calibration; | - | - | - | - | - |
viii. | Reference to the sampling plan and procedures used by the laboratory or other bodies where these are relevant to the validity or application of the results; | - | - | - | - | - |
ix. | The test results with, where appropriate, the units of measurement; | - | - | - | - | - |
x. | The name(s), function(s) and signature(s) or equivalent identification of person(s) authorizing the test report or certificate; and | - | - | - | - | - |
xi. | Where relevant, a statement to the effect that the results relate only to the items tested or calibrated. | - | - | - | - | - |
e. | Material amendments to a test report or calibration certificate after issue shall be made only in the form of a further document, or data transfer, which includes the statement: "Supplement to Test Report (or Calibration Certificate), serial number... (or as otherwise identified)," or an equivalent form of wording. | - | - | - | - | - |
f. | When it is necessary to issue a complete new test report or calibration certificate, this shall be uniquely identified and shall contain a reference to the original that it replaces. | - | - | - | - | - |
g. | If the laboratory chooses to include a reference to their I-502 certification on their test report, any test results not covered under I-502 certification shall be clearly identified on the report. | - | - | - | - | - |
h. | Documentation that the value reported in the CA is within the range and limitations of the analytical method. | - | - | - | - | - |
i. | Documentation that qualitative results (those below the LOQ but above the LOD) are reported as "trace," or with a nonspecific (numerical) designation. | - | - | - | - | - |
j. | Documentation that the methodology has the specificity for the degree of quantitation reported. Final reports are not quantitative to any tenths or hundredths of a percent. | - | - | - | - | - |
k. | Use of appropriate "controls": Documentation of daily use of positive and negative controls that challenge the linearity of the curve; and/or an appropriate "matrix blank" and control with documentation of the performance for each calibration run. | - | - | - | - | - |
29. | Chemical assay procedure/methodology. | - | - | - | - | - |
30. | Quality Control (QC): | - | - | - | - | - |
a. | Documentation of use of an appropriate internal standard for any quantitative measurements as applicable to the method. | - | - | - | - | - |
b. | Appropriate reference standards for quantification of analytes, performing and documenting a calibration curve with each analysis. | - | - | - | - | - |
i. | Reference materials shall, where possible, be traceable to SI units of measurement, or to certified reference materials. Internal reference materials shall be checked for accuracy as far as is technically and economically practicable. | - | - | - | - | - |
ii. | The laboratory shall create and follow procedures for safe handling, transport, storage and use of reference standards and reference materials in order to prevent contamination or deterioration and in order to protect their integrity. | - | - | - | - | - |
iii. | Reference materials shall have a certificate of analysis that documents traceability to a primary standard or certified reference material and associated uncertainty, when possible. When applicable, the certificate must document the specific NIST SRM® or NMI certified reference material used for traceability. | - | - | - | - | - |
c. | Demonstration of calibration curve r2 value of no less than 0.995 with a minimum of four points which bracket the expected sample concentration range. | - | - | - | - | - |
i. | The calibration curve shall be verified by preparing an independently prepared calibration standard (from neat materials) or with a standard from an independent source. Acceptance criteria for the standard calibration curve and the independent calibration verification standard shall be documented. | - | - | - | - | - |
ii. | Instrument calibration/standardization shall be verified each 24-hour period of use, or at each instrument start-up if the instrument is restarted during the 24-hour period, by analysis of a continuing calibration verification standard. Acceptance criteria shall be documented. | - | - | - | - | - |
iii. | Calibration or working quantification ranges shall encompass the concentrations reported by the laboratory. Continuing calibration verification standards and continuing calibration blanks shall be analyzed in accordance with the specified test methods. Acceptance criteria shall be documented. | - | - | - | - | - |
d. | Assuring the quality of test results. | - | - | - | - | - |
i. | The laboratory shall have quality control procedures for monitoring the validity of tests and calibrations undertaken. | - | - | - | - | - |
ii. | The resulting data shall be recorded in such a way that trends are detectable and, where practicable, statistical techniques shall be applied to the reviewing of the results. | - | - | - | - | - |
iii. | This monitoring shall be planned and reviewed and may include, but not be limited to, the following: | - | - | - | - | - |
A. | Regular use of certified reference materials and/or internal quality control using secondary reference materials; | - | - | - | - | - |
B. | Participation in interlaboratory comparison or proficiency-testing programs; | - | - | - | - | - |
C. | Replicate tests or calibrations using the same or different methods; | - | - | - | - | - |
D. | Retesting or recalibration of retained items; | - | - | - | - | - |
E. | Correlation of results for different characteristics of an item. | - | - | - | - | - |
iv. | Quality control data shall be analyzed and, where they are found to be outside predefined criteria, planned actions shall be taken to correct the problem and to prevent incorrect results from occurring. | - | - | - | - | - |
v. | The laboratory shall determine, where feasible, the accuracy and precision of all analyses performed. | - | - | - | - | - |
vi. | Acceptance limits for each method shall be established based on statistical evaluation of the data generated by the analysis of quality control check samples, unless specific acceptance limits are established by the method. | - | - | - | - | - |
vii. | Control charts or quality control databases shall be used to record quality control data and compare them with acceptance limits. | - | - | - | - | - |
viii. | Procedures shall be used to monitor trends and the validity of test results. | - | - | - | - | - |
31. | Proficiency. | - | - | - | - | - |
a. | Participation in approved PT programs for each field of testing. | - | - | - | - | - |
b. | Passing PT results for two consecutive PTs. | - | - | - | - | - |
c. | Documentation of investigation for all failed PTs. | - | - | - | - | - |
32. | Method validation: Scientifically valid testing methodology: AHP monograph compliant, other third-party validation or the current version of a standard method. The following requirements are applied to other third-party validation: | - | - | - | - | - |
a. | The laboratory shall validate nonstandard methods, laboratory-designed/developed methods, standard methods used outside their intended scope, and amplifications and modifications of standard methods to confirm that the methods are fit for the intended use. | - | - | - | - | - |
b. | The validation shall be as extensive as is necessary to meet the needs of a given application or field of application. | - | - | - | - | - |
c. | The laboratory shall record the results obtained, the procedure used for the validation, and a statement as to whether the method is fit for the intended use. | - | - | - | - | - |
d. | The customer shall be informed as to the method chosen. | - | - | - | - | - |
e. | The laboratory shall confirm that it can properly operate standard methods before introducing the tests or calibrations. If the standard method changes, the confirmation shall be repeated. | - | - | - | - | - |
f. | Deviation from test and calibration methods shall occur only if the deviation has been documented, technically justified, authorized, and accepted by the customer. | - | - | - | - | - |
g. | Validation shall be documented and include the following elements as applicable: | - | - | - | - | - |
i. | Minimum acceptance criteria; | - | - | - | - | - |
ii. | Analyte specificity; | - | - | - | - | - |
iii. | Linearity; | - | - | - | - | - |
iv. | Range; | - | - | - | - | - |
v. | Accuracy; | - | - | - | - | - |
vi. | Precision; | - | - | - | - | - |
vii. | Detection limit; | - | - | - | - | - |
viii. | Quantification limit; | - | - | - | - | - |
ix. | Stability of samples and reagents interlaboratory precision; | - | - | - | - | - |
x. | Analysis robustness; | - | - | - | - | - |
xi. | Presence of QC samples; | - | - | - | - | - |
xii. | Use of appropriate internal reference standard; | - | - | - | - | - |
xiii. | Daily monitoring of the response of the instrument; | - | - | - | - | - |
h. | Validation shall be performed for matrix extensions for each type of product tested, including data review of recovery for: | - | - | - | - | - |
i. | Solvent-based extract; | - | - | - | - | - |
ii. | CO2 extraction or other "hash oil"; | - | - | - | - | - |
iii. | Extract made with food grade ethanol; | - | - | - | - | - |
iv. | Extract made with food grade glycerin or propylene glycol; | - | - | - | - | - |
v. | Infused liquids; | - | - | - | - | - |
vi. | Infused solids; | - | - | - | - | - |
vii. | Infused topical preparations; | - | - | - | - | - |
viii. | Other oils, butter or fats. | - | - | - | - | - |
33. | Estimation of uncertainty of measurement. | - | - | - | - | - |
a. | Testing laboratories shall have and shall apply procedures for estimating uncertainty of measurement. The laboratory shall at least attempt to identify all the components of uncertainty and make a reasonable estimation, and shall ensure that the form of reporting of the result does not give a wrong impression of the uncertainty. Reasonable estimation shall be based on knowledge of the performance of the method and on the measurement scope and shall make use of, for example, previous experience and validation data. | - | - | - | - | - |
b. | In those cases where a well-recognized test method specifies limits to the values of the major sources of uncertainty of measurement and specifies the form of presentation of calculated results, the laboratory is considered to have satisfied this clause by following the test method and reporting instructions. | - | - | - | - | - |
c. | When estimating the uncertainty of measurement, all uncertainty components which are of importance in the given situation shall be taken into account using appropriate methods of analysis. | - | - | - | - | - |
d. | Sources contributing to the uncertainty include, but are not necessarily limited to, the reference standards and reference materials used, methods and equipment used, environmental conditions, properties and condition of the item being tested or calibrated, and the operator. | - | - | - | - | - |
e. | Test methods are classified as either qualitative or quantitative. Qualitative tests are defined as having nonnumerical results. Although estimation of measurement uncertainty is not needed for these tests, laboratories are expected to have an understanding of the contributors to variability of the results. For quantitative tests, laboratories shall determine measurement uncertainty using appropriate statistical techniques. | - | - | - | - | - |
f. | Laboratories shall make independent estimations of uncertainty for tests performed on samples with significantly different matrices. | - | - | - | - | - |
g. | Laboratories are required to re-estimate measurement uncertainty when changes to their operations are made that may affect sources of uncertainty. | - | - | - | - | - |
h. | When reporting measurement uncertainty, the test report shall include the coverage factor and confidence level used in the estimations (typically k = approximately 2 at the 95% confidence level). | - | - | - | - | - |
34. | Other methods. | - | - | - | - | - |
a. | Validated microbiological methods fit for purpose. | - | - | - | - | - |
b. | Microbial contaminants within limits as directed by WSLCB. | - | - | - | - | - |
c. | Moisture content testing fit for purpose. Scientifically valid testing methodology: AHP monograph compliant, or other third-party validation. | - | - | - | - | - |
d. | Solvent residuals testing fit for purpose; solvent extracted products made with class 3 or other solvents used are not to exceed 500 parts per million (PPM) per one gram of solvent based product and are to be tested. | - | - | - | - | - |
e. | Any other QA/QC methods is proven to be fit for purpose. | - | - | - | - | - |
35. | Laboratory records. | - | - | - | - | - |
a. | Legible and in ink (or computerized system). | - | - | - | - | - |
b. | Signed and dated. | - | - | - | - | - |
c. | Changes initialed and dated. | - | - | - | - | - |
d. | Evidence of periodic review and signed by a management representative. | - | - | - | - | - |
36. | Preventive/corrective action. | - | - | - | - | - |
| The laboratory shall establish a policy and procedure and shall designate appropriate authorities for implementing corrective action when nonconforming work or departures from the policies and procedures in the management system or technical operations are identified. | - | - | - | - | - |
a. | The procedure for corrective action shall start with an investigation to determine the root cause(s) of the problem. | - | - | - | - | - |
b. | Where corrective action is needed, the laboratory shall identify potential corrective actions. It shall select and implement the action(s) most likely to eliminate the problem and to prevent recurrence. | - | - | - | - | - |
c. | The laboratory shall document and implement any required changes resulting from corrective action investigations. | - | - | - | - | - |
d. | Any PT round that leads to the nonproficient status of a laboratory shall be addressed by the corrective action process. | - | - | - | - | - |
e. | The laboratory shall monitor the results to ensure that the corrective actions taken have been effective. | - | - | - | - | - |
f. | When improvement opportunities are identified or if preventive action is required, action plans shall be developed, implemented and monitored to reduce the likelihood of the occurrence of such nonconformities and to take advantage of the opportunities for improvement. | - | - | - | - | - |
37. | Complaints. | - | - | - | - | - |
a. | The laboratory shall have a policy and procedure for the resolution of complaints received from customers or other parties. | - | - | - | - | - |
b. | Records shall be maintained of all complaints and of the investigations and corrective actions taken by the laboratory. | - | - | - | - | - |
c. | Test reports. | - | - | - | - | - |
d. | Each test report or calibration certificate shall include at least the following information, unless otherwise justified: | - | - | - | - | - |
i. | A title (e.g., "Test Report" or "Calibration Certificate"); | - | - | - | - | - |
ii. | The name and address of the laboratory, and the location where the tests and/or calibrations were carried out, if different from the address of the laboratory; | - | - | - | - | - |
iii. | Unique identification of the test report or calibration certificate (such as the serial number), and on each page an identification in order to ensure that the page is recognized as a part of the test report or calibration certificate, and a clear identification of the end of the test report or calibration certificate; | - | - | - | - | - |
iv. | The name and address of the customer; | - | - | - | - | - |
v. | Identification of the method used; | - | - | - | - | - |
vi. | A description of, the condition of, and unambiguous identification of the item(s) tested or calibrated; | - | - | - | - | - |
vii. | The date of receipt of the test or calibration item(s) where this is critical to the validity and application of the results, and the date(s) of performance of the test or calibration; | - | - | - | - | - |
viii. | Reference to the sampling plan and procedures used by the laboratory or other bodies where these are relevant to the validity or application of the results; | - | - | - | - | - |
ix. | The test or calibration results with, where appropriate, the units of measurement; | - | - | - | - | - |
x. | The name(s), function(s) and signature(s) or equivalent identification of person(s) authorizing the test report or calibration certificate; and | - | - | - | - | - |
xi. | Where relevant, a statement to the effect that the results relate only to the items tested or calibrated. | - | - | - | - | - |
38. | Periodic management review and internal audit. | - | - | - | - | - |
a. | Laboratory management shall annually review its quality system and associated procedures to evaluate continued adequacy. This review shall be documented. | - | - | - | - | - |
b. | Periodically and in accordance with a predetermined schedule perform an internal audit of laboratory operations to verify compliance to the GLP checklist. | - | - | - | - | - |