PDFWAC 246-243-080
Radiation survey instruments.
(1) The licensee shall maintain sufficient calibrated and operable radiation survey instruments at each location where radioactive material is present to make physical radiation surveys as required by this part and chapter 246-221 WAC. Instrumentation required by this section shall be capable of measuring a range from 0.02 millisieverts (2 millirems) per hour through 0.01 sievert (1 rem) per hour.
(2) Each radiation survey instrument shall be calibrated:
(a) At intervals not to exceed six months and after each instrument servicing except for battery changes;
(b) Such that accuracy within ± 20 percent of the calibration source can be demonstrated at each point checked; and
(c) For linear scale instruments, at two points located approximately one-third and two-thirds of full scale on each scale; for logarithmic scale instruments, at mid-range of each decade; and for digital instruments at three points between 0.02 and 10 millisieverts (2 and 1000 millirems) per hour.
(3) Records shall be maintained of these calibrations for three years after the calibration date for inspection by the department.
[Statutory Authority: RCW 70.98.050. WSR 00-08-013, § 246-243-080, filed 3/24/00, effective 4/24/00; WSR 94-01-073, § 246-243-080, filed 12/9/93, effective 1/9/94. Statutory Authority: RCW 70.98.050 and 70.98.080. WSR 91-15-112 (Order 184), § 246-243-080, filed 7/24/91, effective 8/24/91. Statutory Authority: RCW 43.70.040. WSR 91-02-049 (Order 121), recodified as § 246-243-080, filed 12/27/90, effective 1/31/91. Statutory Authority: RCW 70.98.080. WSR 83-19-050 (Order 2026), § 402-36-060, filed 9/16/83. Statutory Authority: RCW 70.98.050. WSR 81-01-011 (Order 1570), § 402-36-060, filed 12/8/80; Order 1084, § 402-36-060, filed 1/14/76; Order 1, § 402-36-060, filed 1/8/69; Rules (part), filed 10/26/66.]