HTML has links - PDF has Authentication
296-62-07447  <<  296-62-07449 >>   296-62-07451

PDFWAC 296-62-07449

Appendix E—Cadmium in workplace atmospheres.

Method number: ID-189 (OSHA); (ICP/MS) 0009 (WISHA)
Matrix: Air
WISHA permissible exposure limits: 5 µg/m3 (TWA), 2.5 µg/m3 (action level TWA)
Collection procedure: A known volume of air is drawn through a 37-mm diameter filter cassette containing a 0.8 µm mixed cellulose ester membrane filter (MCEF).
Recommended air volume: 960 L
Recommended sampling rate: 2.0 L/min
Analytical procedure: Air filter samples are digested with nitric acid. After digestion, a small amount of hydrochloric acid is added. The samples are then diluted to volume with deionized water and analyzed by either flame atomic absorption spectroscopy (AAS) or flameless atomic absorption spectroscopy using a heated graphite furnace atomizer (AAS-HGA).
Detection limits:
Qualitative: 0.2 µg/m3 for a 200 L sample by Flame AAS, 0.007 µg/m3 for a 60 L sample by AAS-HGA
Quantitative: 0.70 µg/m3 for a 200 L sample by Flame AAS, 0.025 µg/m3 for a 60 L sample by AAS-HGA
Precision and accuracy: (Flame AAS Analysis and AAS-HGA Analysis):
Validation level: 2.5 to 10 µg/m3 for a 400 L air vol, 1.25 to 5.0 µg/m3 for a 60 L air vol CV1 (pooled): 0.010, 0.043
Analytical bias: +4.0%, -5.8%
Overall analytical error: ±6.0%, ±14.2%
Method classification: Validated Date: June, 1992
Inorganic Service Branch II, OSHA Salt Lake Technical Center, Salt Lake City, Utah Commercial manufacturers and products mentioned in this method are for descriptive use only and do not constitute endorsements by USDOL-OSHA. Similar products from other sources can be substituted.
(1) Introduction.
(a) Scope.
This method describes the collection of airborne elemental cadmium and cadmium compounds on 0.8 µm mixed cellulose ester membrane filters and their subsequent analysis by either flame atomic absorption spectroscopy (AAS) or flameless atomic absorption spectroscopy using a heated graphite furnace atomizer (AAS-HGA). It is applicable for both TWA and action level TWA permissible exposure level (PEL) measurements. The two atomic absorption analytical techniques included in the method do not differentiate between cadmium fume and cadmium dust samples. They also do not differentiate between elemental cadmium and its compounds.
(b) Principle.
Airborne elemental cadmium and cadmium compounds are collected on a 0.8 µm mixed cellulose ester membrane filter (MCEF). The air filter samples are digested with concentrated nitric acid to destroy the organic matrix and dissolve the cadmium analytes. After digestion, a small amount of concentrated hydrochloric acid is added to help dissolve other metals which may be present. The samples are diluted to volume with deionized water and then aspirated into the oxidizing air/acetylene flame of an atomic absorption spectrophotometer for analysis of elemental cadmium. If the concentration of cadmium in a sample solution is too low for quantitation by this flame AAS analytical technique, and the sample is to be averaged with other samples for TWA calculations, aliquots of the sample and a matrix modifier are later injected onto a L'vov platform in a pyrolytically-coated graphite tube of a Zeeman atomic absorption spectrophotometer/graphite furnace assembly for analysis of elemental cadmium. The matrix modifier is added to stabilize the cadmium metal and minimize sodium chloride as an interference during the high temperature charring step of the analysis subsection (5)(a) and (b) of this section.
(c) History.
Previously, two OSHA sampling and analytical methods for cadmium were used concurrently WAC 296-62-07449 (5)(c) and (d). Both of these methods also required 0.8 µm mixed cellulose ester membrane filters for the collection of air samples. These cadmium air filter samples were analyzed by either flame atomic absorption spectroscopy (subsection (5)(c) of this section) or inductively coupled plasma/atomic emission spectroscopy (ICP-AES) (subsection (5)(d) of this section). Neither of these two analytical methods have adequate sensitivity for measuring workplace exposure to airborne cadmium at the new lower TWA and action level TWA PEL levels when consecutive samples are taken on one employee and the sample results need to be averaged with other samples to determine a single TWA. The inclusion of two atomic absorption analytical techniques in the new sampling and analysis method for airborne cadmium permits quantitation of sample results over a broad range of exposure levels and sampling periods. The flame AAS analytical technique included in this method is similar to the previous procedure given in the General Metals Method ID-121 (subsection (5)(c) of this section) with some modifications. The sensitivity of the AAS-HGA analytical technique included in this method is adequate to measure exposure levels at 1/10 the action level TWA, or lower, when less than full-shift samples need to be averaged together.
(d) Properties (subsection (5)(e) of this section).
Elemental cadmium is a silver-white, blue-tinged, lustrous metal which is easily cut with a knife. It is slowly oxidized by moist air to form cadmium oxide. It is insoluble in water, but reacts readily with dilute nitric acid. Some of the physical properties and other descriptive information of elemental cadmium are given below:
CAS No . . . .
7440-43-9
Atomic Number . . . .
48
Atomic Symbol . . . .
Cd
Atomic Weight . . . .
112.41
Melting Point . . . .
321°C
Boiling Point . . . .
765°C
Density . . . .
8.65 g/mL (25°C)
The properties of specific cadmium compounds are described in reference subsection (5)(e) of this section.
(e) Method performance.
A synopsis of method performance is presented below. Further information can be found in subsection (4) of this section.
(i) The qualitative and quantitative detection limits for the flame AAS analytical technique are 0.04 µg (0.004 µg/mL) and 0.14 µg (0.014 µg/mL) cadmium, respectively, for a 10 mL solution volume. These correspond, respectively, to 0.2 µg/m3 and 0.70 µg/m3 for a 200 L air volume.
(ii) The qualitative and quantitative detection limits for the AAS-HGA analytical technique are 0.44 ng (0.044 ng/mL) and 1.5 ng (0.15 ng/mL) cadmium, respectively, for a 10 mL solution volume. These correspond, respectively, to 0.007 µg/m3 and 0.025 µg/m3 for a 60 L air volume.
(iii) The average recovery by the flame AAS analytical technique of 17 spiked MCEF samples containing cadmium in the range of 0.5 to 2.0 times the TWA target concentration of 5 µg/m3 (assuming a 400 L air volume) was 104.0% with a pooled coefficient of variation (CV1) of 0.010. The flame analytical technique exhibited a positive bias of +4.0% for the validated concentration range. The overall analytical error (OAE) for the flame AAS analytical technique was ±6.0%.
(iv) The average recovery by the AAS-HGA analytical technique of 18 spiked MCEF samples containing cadmium in the range of 0.5 to 2.0 times the action level TWA target concentration of 2.5 µg/m3 (assuming a 60 L air volume) was 94.2% with a pooled coefficient of variation (CV1) of 0.043. The AAS-HGA analytical technique exhibited a negative bias of -5.8% for the validated concentration range. The overall analytical error (OAE) for the AAS-HGA analytical technique was ±14.2%.
(v) Sensitivity in flame atomic absorption is defined as the characteristic concentration of an element required to produce a signal of 1% absorbance (0.0044 absorbance units). Sensitivity values are listed for each element by the atomic absorption spectrophotometer manufacturer and have proved to be a very valuable diagnostic tool to determine if instrumental parameters are optimized and if the instrument is performing up to specification. The sensitivity of the spectrophotometer used in the validation of the flame AAS analytical technique agreed with the manufacturer specifications (subsection (5)(f) of this section); the 2 µg/mL cadmium standard gave an absorbance reading of 0.350 abs. units.
(vi) Sensitivity in graphite furnace atomic absorption is defined in terms of the characteristic mass, the number of picograms required to give an integrated absorbance value of 0.0044 absorbance-second (subsection (5)(g) of this section). Data suggests that under stabilized temperature platform furnace (STPF) conditions (see (f)(ii) of this subsection), characteristic mass values are transferable between properly functioning instruments to an accuracy of about twenty percent (subsection (5)(b) of this section). The characteristic mass for STPF analysis of cadmium with Zeeman background correction listed by the manufacturer of the instrument used in the validation of the AAS-HGA analytical technique was 0.35 pg. The experimental characteristic mass value observed during the determination of the working range and detection limits of the AAS-HGA analytical technique was 0.41 pg.
(f) Interferences.
(i) High concentrations of silicate interfere in determining cadmium by flame AAS (subsection (5)(f) of this section). However, silicates are not significantly soluble in the acid matrix used to prepare the samples.
(ii) Interferences, such as background absorption, are reduced to a minimum in the AAS-HGA analytical technique by taking full advantage of the stabilized temperature platform furnace (STPF) concept. STPF includes all of the following parameters (subsection (5)(b) of this section):
(A) Integrated absorbance;
(B) Fast instrument electronics and sampling frequency;
(C) Background correction;
(D) Maximum power heating;
(E) Atomization off the L'vov platform in a pyrolytically coated graphite tube;
(F) Gas stop during atomization;
(G) Use of matrix modifiers.
(g) Toxicology (subsection (5)(n) of this section).
Information listed within this section is synopsis of current knowledge of the physiological effects of cadmium and is not intended to be used as the basis for WISHA policy. IARC classifies cadmium and certain of its compounds as Group 2A carcinogens (probably carcinogenic to humans). Cadmium fume is intensely irritating to the respiratory tract. Workplace exposure to cadmium can cause both chronic and acute effects. Acute effects include tracheobronchitis, pneumonitis, and pulmonary edema. Chronic effects include anemia, rhinitis/anosmia, pulmonary emphysema, proteinuria and lung cancer. The primary target organs for chronic disease are the kidneys (noncarcinogenic) and the lungs (carcinogenic).
(2) Sampling.
(a) Apparatus.
(i) Filter cassette unit for air sampling: A 37-mm diameter mixed cellulose ester membrane filter with a pore size of 0.8 µm contained in a 37-mm polystyrene two- or three-piece cassette filter holder (part no. MAWP 037 A0, Millipore Corp., Bedford, MA). The filter is supported with a cellulose backup pad. The cassette is sealed prior to use with a shrinkable gel band.
(ii) A calibrated personal sampling pump whose flow is determined to an accuracy of ±5% at the recommended flow rate with the filter cassette unit in line.
(b) Procedure
(i) Attach the prepared cassette to the calibrated sampling pump (the backup pad should face the pump) using flexible tubing. Place the sampling device on the employee such that air is sampled from the breathing zone.
(ii) Collect air samples at a flow rate of 2.0 L/min. If the filter does not become overloaded, a full-shift (at least seven hours) sample is strongly recommended for TWA and action level TWA measurements with a maximum air volume of 960 L. If overloading occurs, collect consecutive air samples for shorter sampling periods to cover the full workshift.
(iii) Replace the end plugs into the filter cassettes immediately after sampling. Record the sampling conditions.
(iv) Securely wrap each sample filter cassette end-to-end with a sample seal.
(v) Submit at least one blank sample. With each set of air samples. The blank sample should be handled the same as the other samples except that no air is drawn through it.
(vi) Ship the samples to the laboratory for analysis as soon as possible in a suitable container designed to prevent damage in transit.
(3) Analysis.
(a) Safety precautions.
(i) Wear safety glasses, protective clothing and gloves at all times.
(ii) Handle acid solutions with care. Handle all cadmium samples and solutions with extra care (see subsection (1)(g) of this section). Avoid their direct contact with work area surfaces, eyes, skin and clothes. Flush acid solutions which contact the skin or eyes with copious amounts of water.
(iii) Perform all acid digestions and acid dilutions in an exhaust hood while wearing a face shield. To avoid exposure to acid vapors, do not remove beakers containing concentrated acid solutions from the exhaust hood until they have returned to room temperature and have been diluted or emptied.
(iv) Exercise care when using laboratory glassware. Do not use chipped pipets, volumetric flasks, beakers or any glassware with sharp edges exposed in order to avoid the possibility of cuts or abrasions.
(v) Never pipet by mouth.
(vi) Refer to the instrument instruction manuals and SOPs (subsection (5)(h) and (i) of this section) for proper and safe operation of the atomic absorption spectrophotometer, graphite furnace atomizer and associated equipment.
(vii) Because metallic elements and other toxic substances are vaporized during AAS flame or graphite furnace atomizer operation, it is imperative that an exhaust vent be used. Always ensure that the exhaust system is operating properly during instrument use.
(b) Apparatus for sample and standard preparation.
(i) Hot plate, capable of reaching 150°C, installed in an exhaust hood.
(ii) Phillips beakers, 125 mL.
(iii) Bottles, narrow-mouth, polyethylene or glass with leakproof caps: used for storage of standards and matrix modifier.
(iv) Volumetric flasks, volumetric pipets, beakers and other associated general laboratory glassware.
(v) Forceps and other associated general laboratory equipment.
(c) Apparatus for flame AAS analysis.
(i) Atomic absorption spectrophotometer consisting of a(an):
Nebulizer and burner head; pressure regulating devices capable of maintaining constant oxidant and fuel pressures; optical system capable of isolating the desired wavelength of radiation (228.8 nm); adjustable slit; light measuring and amplifying device; display, strip chart, or computer interface for indicating the amount of absorbed radiation; cadmium hollow cathode lamp or electrodeless discharge lamp (EDL) and power supply.
(ii) Oxidant: Compressed air, filtered to remove water, oil and other foreign substances.
(iii) Fuel: Standard commercially available tanks of acetylene dissolved in acetone; tanks should be equipped with flash arresters.
Caution:
Do not use grades of acetylene containing solvents other than acetone because they may damage the PVC tubing used in some instruments.
(iv) Pressure-reducing valves: Two gauge, two-stage pressure regulators to maintain fuel and oxidant pressures somewhat higher than the controlled operating pressures of the instrument.
(v) Exhaust vent installed directly above the spectrophotometer burner head.
(d) Apparatus for AAS-HGA analysis.
(i) Atomic absorption spectrophotometer consisting of a(an):
Heated graphite furnace atomizer (HGA) with argon purge system pressure-regulating devices capable of maintaining constant argon purge pressure; optical system capable of isolating the desired wavelength of radiation (228.8 nm); adjustable slit; light measuring and amplifying device; display, strip chart, or computer interface for indicating the amount of absorbed radiation (as integrated absorbance, peak area); background corrector: Zeeman or deuterium arc. The Zeeman background corrector is recommended; cadmium hollow cathode lamp or electrodeless discharge lamp (EDL) and power supply; autosampler capable of accurately injecting 5 to 20 µL sample aliquots onto the L'vov Platform in a graphite tube.
(ii) Pyrolytically coated graphite tubes containing solid, pyrolytic L'vov platforms.
(iii) Polyethylene sample cups, 2.0 to 2.5 mL, for use with the autosampler.
(iv) Inert purge gas for graphite furnace atomizer: Compressed gas cylinder of purified argon.
(v) Two gauge, two-stage pressure regulator for the argon gas cylinder.
(vi) Cooling water supply for graphite furnace atomizer.
(vii) Exhaust vent installed directly above the graphite furnace atomizer.
(e) Reagents. All reagents should be ACS analytical reagent grade or better.
(i) Deionized water with a specific conductance of less than 10 µS.
(ii) Concentrated nitric acid, HNO3.
(iii) Concentrated hydrochloric acid, HCl.
(iv) Ammonium phosphate, monobasic, NH4H2PO4.
(v) Magnesium nitrate, Mg(NO3)2 • 6H2O.
(vi) Diluting solution (4% HNO3, 0.4% HCl): Add 40 mL HNO3 and 4 mL HCl carefully to approximately 500 mL deionized water and dilute to 1 L with deionized water.
(vii) Cadmium standard stock solution, 1,000 µg/mL: Use a commercially available certified 1,000 µg/mL cadmium standard or, alternatively, dissolve 1.0000 g of cadmium metal in a minimum volume of 1:1 HCl and dilute to 1 L with 4% HNO3. Observe expiration dates of commercial standards. Properly dispose of commercial standards with no expiration dates or prepared standards one year after their receipt or preparation date.
(viii) Matrix modifier for AAS-HGA analysis: Dissolve 1.0 g NH4H2PO4 and 0.15 g Mg(NO3)2 • 6H2O in approximately 200 mL deionized water. Add 1 mL HNO3 and dilute to 500 mL with deionized water.
(ix) Nitric Acid, 1:1 HNO3/DI H2O mixture: Carefully add a measured volume of concentrated HNO3 to an equal volume of DI H2O.
(x) Nitric acid, 10% v/v: Carefully add 100 mL of concentrated HNO3 to 500 mL of DI H2O and dilute to 1 L.
(f) Glassware preparation.
(i) Clean Phillips beakers by refluxing with 1:1 nitric acid on a hot plate in a fume hood. Thoroughly rinse with deionized water and invert the beakers to allow them to drain dry.
(ii) Rinse volumetric flasks and all other glassware with 10% nitric acid and deionized water prior to use.
(g) Standard preparation for flame AAS analysis.
(i) Dilute stock solutions: Prepare 1, 5, 10 and 100 µg/mL cadmium standard stock solutions by making appropriate serial dilutions of 1,000 µg/mL cadmium standard stock solution with the diluting solution described in (e)(vi) of this subsection.
(ii) Working standards: Prepare cadmium working standards in the range of 0.02 to 2.0 µg/mL by making appropriate serial dilutions of the dilute stock solutions with the same diluting solution. A suggested method of preparation of the working standards is given below.
Working
standard
(µg/mL)
Std
solution
(µg/mL)
Aliquot
(mL)
Final vol.
(mL)
0.02
1
10
500
0.05
5
5
500
0.1
10
5
500
0.2
10
10
500
0.5
10
25
500
1
100
5
500
2
100
10
500
Store the working standards in 500-mL, narrow-mouth polyethylene or glass bottles with leak proof caps. Prepare every twelve months.
(h) Standard preparation for AAS-HGA analysis.
(i) Dilute stock solutions: Prepare 10, 100 and 1,000 ng/mL cadmium standard stock solutions by making appropriate ten-fold serial dilutions of the 1,000 µg/mL cadmium standard stock solution with the diluting solution described in (e)(vi) of this subsection.
(ii) Working standards: Prepare cadmium working standards in the range of 0.2 to 20 ng/mL by making appropriate serial dilutions of the dilute stock solutions with the same diluting solution. A suggested method of preparation of the working standards is given below.
Working
standard
(ng/mL)
Std
solution
(ng/mL)
Aliquot
(mL)
Final vol.
(mL)
0.2
10
2
100
0.5
10
5
100
1
10
10
100
2
100
2
100
5
100
5
100
10
100
10
100
20
1,000
2
100
Store the working standards in narrow-mouth polyethylene or glass bottles with leakproof caps. Prepare monthly.
(i) Sample preparation.
(i) Carefully transfer each sample filter with forceps from its filter cassette unit to a clean, separate 125-mL Phillips beaker along with any loose dust found in the cassette. Label each Phillips beaker with the appropriate sample number.
(ii) Digest the sample by adding 5 mL of concentrated nitric acid (HNO3) to each Phillips beaker containing an air filter sample. Place the Phillips beakers on a hot plate in an exhaust hood and heat the samples until approximately 0.5 mL remains. The sample solution in each Phillips beaker should become clear. If it is not clear, digest the sample with another portion of concentrated nitric acid.
(iii) After completing the HNO3 digestion and cooling the samples, add 40 µL (2 drops) of concentrated HCl to each air sample solution and then swirl the contents. Carefully add about 5 mL of deionized water by pouring it down the inside of each beaker.
(iv) Quantitatively transfer each cooled air sample solution from each Phillips beaker to a clean 10-mL volumetric flask. Dilute each flask to volume with deionized water and mix well.
(j) Flame AAS analysis.
Analyze all of the air samples for their cadmium content by flame atomic absorption spectroscopy (AAS) according to the instructions given below.
(i) Set up the atomic absorption spectrophotometer for the air/acetylene flame analysis of cadmium according to the SOP (subsection (5)(h) of this section) or the manufacturer's operational instructions. For the source lamp, use the cadmium hollow cathode or electrodeless discharge lamp operated at the manufacturer's recommended rating for continuous operation. Allow the lamp to warm up ten to twenty minutes or until the energy output stabilizes. Optimize conditions such as lamp position, burner head alignment, fuel and oxidant flow rates, etc. See the SOP or specific instrument manuals for details. Instrumental parameters for the Perkin-Elmer Model 603 used in the validation of this method are given in subsection (6) of this section.
(ii) Aspirate and measure the absorbance of a standard solution of cadmium. The standard concentration should be within the linear range. For the instrumentation used in the validation of this method a 2 µg/mL cadmium standard gives a net absorbance reading of about 0.350 abs. units (see subsection (1)(e)(v) of this section) when the instrument and the source lamp are performing to manufacturer specifications.
(iii) To increase instrument response, scale expand the absorbance reading of the aspirated 2 µg/mL working standard approximately four times. Increase the integration time to at least three seconds to reduce signal noise.
(iv) Autozero the instrument while aspirating a deionized water blank. Monitor the variation in the baseline absorbance reading (baseline noise) for a few minutes to insure that the instrument, source lamp and associated equipment are in good operating condition.
(v) Aspirate the working standards and samples directly into the flame and record their absorbance readings. Aspirate the deionized water blank immediately after every standard or sample to correct for and monitor any baseline drift and noise. Record the baseline absorbance reading of each deionized water blank. Label each standard and sample reading and its accompanying baseline reading.
(vi) It is recommended that the entire series of working standards be analyzed at the beginning and end of the analysis of a set of samples to establish a concentration-response curve, ensure that the standard readings agree with each other and are reproducible. Also, analyze a working standard after every five or six samples to monitor the performance of the spectrophotometer. Standard readings should agree within ±10 to 15% of the readings obtained at the beginning of the analysis.
(vii) Bracket the sample readings with standards during the analysis. If the absorbance reading of a sample is above the absorbance reading of the highest working standard, dilute the sample with diluting solution and reanalyze. Use the appropriate dilution factor in the calculations.
(viii) Repeat the analysis of approximately ten percent of the samples for a check of precision.
(ix) If possible, analyze quality control samples from an independent source as a check on analytical recovery and precision.
(x) Record the final instrument settings at the end of the analysis. Date and label the output.
(k) AAS-HGA analysis.
Initially analyze all of the air samples for their cadmium content by flame atomic absorption spectroscopy (AAS) according to the instructions given in (j) of this subsection. If the concentration of cadmium in a sample solution is less than three times the quantitative detection limit (0.04 µg/mL (40 ng/mL) for the instrumentation used in the validation) and the sample results are to be averaged with other samples for TWA calculations, proceed with the AAS-HGA analysis of the sample as described below.
(i) Set up the atomic absorption spectrophotometer and HGA for flameless atomic absorption analysis of cadmium according to the SOP (subsection (5)(i) of this section) or the manufacturer's operational instructions and allow the instrument to stabilize. The graphite furnace atomizer is equipped with a pyrolytically coated graphite tube containing a pyrolytic platform. For the source lamp, use a cadmium hollow cathode or electrodeless discharge lamp operated at the manufacturer's recommended setting for graphite furnace operation. The Zeeman background corrector and EDL are recommended for use with the L'vov platform. Instrumental parameters for the Perkin-Elmer Model 5100 spectrophotometer and Zeeman HGA-600 graphite furnace used in the validation of this method are given in subsection (7) of this section.
(ii) Optimize the energy reading of the spectrophotometer at 228.8 nm by adjusting the lamp position and the wavelength according to the manufacturer's instructions.
(iii) Set up the autosampler to inject a 5-µL aliquot of the working standard, sample or reagent blank solution onto the L'vov platform along with a 10-µL overlay of the matrix modifier.
(iv) Analyze the reagent blank (diluting solution, (e)(vi) of this subsection) and then autozero the instrument before starting the analysis of a set of samples. It is recommended that the reagent blank be analyzed several times during the analysis to assure the integrated absorbance (peak area) reading remains at or near zero.
(v) Analyze a working standard approximately midway in the linear portion of the working standard range two or three times to check for reproducibility and sensitivity (see subsection (1)(e)(v) and (vi) of this section) before starting the analysis of samples. Calculate the experimental characteristic mass value from the average integrated absorbance reading and injection volume of the analyzed working standard. Compare this value to the manufacturer's suggested value as a check of proper instrument operation.
(vi) Analyze the reagent blank, working standard, and sample solutions. Record and label the peak area (abs-sec) readings and the peak and background peak profiles on the printer/plotter.
(vii) It is recommended the entire series of working standards be analyzed at the beginning and end of the analysis of a set of samples. Establish a concentration-response curve and ensure standard readings agree with each other and are reproducible. Also, analyze a working standard after every five or six samples to monitor the performance of the system. Standard readings should agree within ±15% of the readings obtained at the beginning of the analysis.
(viii) Bracket the sample readings with standards during the analysis. If the peak area reading of a sample is above the peak area reading of the highest working standard, dilute the sample with the diluting solution and reanalyze. Use the appropriate dilution factor in the calculations.
(ix) Repeat the analysis of approximately ten percent of the samples for a check of precision.
(x) If possible, analyze quality control samples from an independent source as a check of analytical recovery and precision.
(xi) Record the final instrument settings at the end of the analysis. Date and label the output.
(l) Calculations.
Note:
Standards used for HGA analysis are in ng/mL. Total amounts of cadmium from calculations will be in ng (not µg) unless a prior conversion is made.
(i) Correct for baseline drift and noise in flame AAS analysis by subtracting each baseline absorbance reading from its corresponding working standard or sample absorbance reading to obtain the net absorbance reading for each standard and sample.
(ii) Use a least squares regression program to plot a concentration-response curve of net absorbance reading (or peak area for HGA analysis) versus concentration (µg/mL or ng/mL) of cadmium in each working standard.
(iii) Determine the concentration (µg/mL or ng/mL) of cadmium in each sample from the resulting concentration-response curve. If the concentration of cadmium in a sample solution is less than three times the quantitative detection limit (0.04 µg/mL (40 ng/mL) for the instrumentation used in the validation of the method) and if consecutive samples were taken on one employee and the sample results are to be averaged with other samples to determine a single TWA, reanalyze the sample by AAS-HGA as described in (k) of this subsection and report the AAS-HGA analytical results.
(iv) Calculate the total amount (µg or ng) of cadmium in each sample from the sample solution volume (mL):
W=(C)(sample vol, mL)(DF)
Where:
W=Total cadmium in sample
 
C=Calculated concentration of cadmium
 
DF=Dilution Factor (if applicable)
(v) Make a blank correction for each air sample by subtracting the total amount of cadmium in the corresponding blank sample from the total amount of cadmium in the sample.
(vi) Calculate the concentration of cadmium in an air sample (mg/m3 or µg/m3) by using one of the following equations:
mg/m3=Wbc/(Air vol sampled, L)
 
or
 
µg/m3=(Wbc)(1,000 ng/µg)/(Air vol sampled, L)
Where:
Wbc=blank corrected total µg cadmium in the sample. (1µg=1,000 ng)
(4) Backup data.
(a) Introduction.
(i) The purpose of this evaluation is to determine the analytical method recovery, working standard range, and qualitative and quantitative detection limits of the two atomic absorption analytical techniques included in this method. The evaluation consisted of the following experiments:
(A) An analysis of twenty-four samples (six samples each at 0.1, 0.5, 1 and 2 times the TWA-PEL) for the analytical method recovery study of the flame AAS analytical technique.
(B) An analysis of eighteen samples (six samples each at 0.5, 1 and 2 times the action level TWA-PEL) for the analytical method recovery study of the AAS-HGA analytical technique.
(C) Multiple analyses of the reagent blank and a series of standard solutions to determine the working standard range and the qualitative and quantitative detection limits for both atomic absorption analytical techniques.
(ii) The analytical method recovery results at all test levels were calculated from concentration-response curves and statistically examined for outliers at the ninety-nine percent confidence level. Possible outliers were determined using the Treatment of Outliers test (subsection (5)(j) of this section). In addition, the sample results of the two analytical techniques, at 0.5, 1.0 and 2.0 times their target concentrations, were tested for homogeneity of variances also at the ninety-nine percent confidence level. Homogeneity of the coefficients of variation was determined using the Bartlett's test (subsection (5)(k) of this section). The overall analytical error (OAE) at the ninety-five percent confidence level was calculated using the equation (subsection (5)(l) of this section):
OAE=±[│Bias│+(1.96)(CV1(pooled))(100%)]
(iii) A derivation of the International Union of Pure and Applied Chemistry (IUPAC) detection limit equation (subsection (5)(m) of this section) was used to determine the qualitative and quantitative detection limits for both atomic absorption analytical techniques:
Cld=k(sd)/m
(Equation 1)
 
Where:
Cld=the smallest reliable detectable concentration an analytical instrument can determine at a given confidence level.
 
k=3 for the Qualitative Detection Limit at the 99.86% Confidence Level
 
=10 for the Quantitative Detection Limit at the 99.99% Confidence Level.
 
sd=standard deviation of the reagent blank (Rbl) readings.
 
m=analytical sensitivity or slope as calculated by linear regression.
(iv) Collection efficiencies of metallic fume and dust atmospheres on 0.8-µm mixed cellulose ester membrane filters are well documented and have been shown to be excellent (subsection (5)(k) of this section). Since elemental cadmium and the cadmium component of cadmium compounds are nonvolatile, stability studies of cadmium spiked MCEF samples were not performed.
(b) Equipment.
(i) A Perkin-Elmer (PE) Model 603 spectrophotometer equipped with a manual gas control system, a stainless steel nebulizer, a burner mixing chamber, a flow spoiler and a 10 cm (one-slot) burner head was used in the experimental validation of the flame AAS analytical technique. A PE cadmium hollow cathode lamp, operated at the manufacturer's recommended current setting for continuous operation (4 mA), was used as the source lamp. Instrument parameters are listed in subsection (6) of this section.
(ii) A PE Model 5100 spectrophotometer, Zeeman HGA-600 graphite furnace atomizer and AS-60 HGA autosampler were used in the experimental validation of the AAS-HGA analytical technique. The spectrophotometer was equipped with a PE Series 7700 professional computer and Model PR-310 printer. A PE System 2 cadmium electrodeless discharge lamp, operated at the manufacturer's recommended current setting for modulated operation (170 mA), was used as the source lamp. Instrument parameters are listed in subsection (7) of this section.
(c) Reagents.
(i) J.T. Baker Chem. Co. (Analyzed grade) concentrated nitric acid, 69.0-71.0%, and concentrated hydrochloric acid, 36.5-38.0%, were used to prepare the samples and standards.
(ii) Ammonium phosphate, monobasic, NH4H2PO4 and magnesium nitrate hexahydrate, Mg(NO3)2.6 H2O both manufactured by the Mallinckrodt Chem. Co., were used to prepare the matrix modifier for AAS-HGA analysis.
(d) Standard preparation for flame AAS analysis.
(i) Dilute stock solutions: Prepared 0.01, 0.1, 1, 10 and 100 µg/mL cadmium standard stock solutions by making appropriate serial dilutions of a commercially available 1,000 µg/mL cadmium standard stock solution (RICCA Chemical Co., Lot# A102) with the diluting solution (4% HNO3, 0.4% HCl).
(ii) Analyzed standards: Prepared cadmium standards in the range of 0.001 to 2.0 µg/mL by pipetting 2 to 10 mL of the appropriate dilute cadmium stock solution into a 100-mL volumetric flask and diluting to volume with the diluting solution. (See subsection (3)(g)(ii) of this section).
(e) Standard preparation for AAS-HGA analysis.
(i) Dilute stock solutions: Prepared 1, 10, 100 and 1,000 ng/mL cadmium standard stock solutions by making appropriate serial dilutions of a commercially available 1,000 µg/mL cadmium standard stock solution (J.T. Baker Chemical Co., Instra-analyzed, Lot# D22642) with the diluting solution (4% HNO3, 0.4% HCl).
(ii) Analyzed standards: Prepared cadmium standards in the range of 0.1 to 40 ng/mL by pipetting 2 to 10 mL of the appropriate dilute cadmium stock solution into a 100-mL volumetric flask and diluting to volume with the diluting solution. (See subsection (3)(h)(ii) of this section).
(f) Detection limits and standard working range for flame AAS analysis.
(i) Analyzed the reagent blank solution and the entire series of cadmium standards in the range of 0.001 to 2.0 µg/mL three to six times according to the instructions given in subsection (3)(j) of this section. The diluting solution (4% HNO3, 0.4% HCl) was used as the reagent blank. The integration time on the PE 603 spectrophotometer was set to 3.0 seconds and a four-fold expansion of the absorbance reading of the 2.0 µg/mL cadmium standard was made prior to analysis. The 2.0 µg/mL standard gave a net absorbance reading of 0.350 abs. units prior to expansion in agreement with the manufacturer's specifications (subsection (5)(f) of this section).
(ii) The net absorbance readings of the reagent blank and the low concentration Cd standards from 0.001 to 0.1 µg/mL and the statistical analysis of the results are shown in Table 1. The standard deviation, sd, of the six net absorbance readings of the reagent blank is 1.05 abs. units. The slope, m, as calculated by a linear regression plot of the net absorbance readings (shown in Table 2) of the 0.02 to 1.0 µg/mL cadmium standards versus their concentration is 772.7 abs. units/(µg/mL).
(iii) If these values for sd and the slope, m, are used in Eqn. 1 ((a)(ii) of this subsection), the qualitative and quantitative detection limits as determined by the IUPAC Method are:
Cld=
(3)(1.05 abs. units)/(772.7 abs. units/(µg/mL))= 0.0041 µg/mL for the qualitative detection limit.
Cld=
(10)(1.05 abs. units)/(772.7 abs. units/µg/mL)) =0.014 µg/mL for the quantitative detection limit.
The qualitative and quantitative detection limits for the flame AAS analytical technique are 0.041 µg and 0.14 µg cadmium, respectively, for a 10 mL solution volume. These correspond, respectively, to 0.2 µg/m3 and 0.70 µg/m3 for a 200 L air volume.
(iv) The recommended Cd standard working range for flame AAS analysis is 0.02 to 2.0 µg/mL. The net absorbance readings of the reagent blank and the recommended working range standards and the statistical analysis of the results are shown in Table 2. The standard of lowest concentration in the working range, 0.02 µg/mL, is slightly greater than the calculated quantitative detection limit, 0.014 µg/mL. The standard of highest concentration in the working range, 2.0 µg/mL, is at the upper end of the linear working range suggested by the manufacturer (subsection (5)(f) of this section). Although the standard net absorbance readings are not strictly linear at concentrations above 0.5 µg/mL, the deviation from linearity is only about ten percent at the upper end of the recommended standard working range. The deviation from linearity is probably caused by the four-fold expansion of the signal suggested in the method. As shown in Table 2, the precision of the standard net absorbance readings are excellent throughout the recommended working range; the relative standard deviations of the readings range from 0.009 to 0.064.
(g) Detection limits and standard working range for AAS-HGA analysis.
(i) Analyzed the reagent blank solution and the entire series of cadmium standards in the range of 0.1 to 40 ng/mL according to the instructions given in subsection (3)(k) of this section. The diluting solution (4% HNO3, 0.4% HCl) was used as the reagent blank. A fresh aliquot of the reagent blank and of each standard was used for every analysis. The experimental characteristic mass value was 0.41 pg, calculated from the average peak area (abs-sec) reading of the 5 ng/mL standard which is approximately midway in the linear portion of the working standard range. This agreed within twenty percent with the characteristic mass value, 0.35 pg, listed by the manufacturer of the instrument (subsection (5)(b) of this section).
(ii) The peak area (abs-sec) readings of the reagent blank and the low concentration Cd standards from 0.1 to 2.0 ng/mL and statistical analysis of the results are shown in Table 3. Five of the reagent blank peak area readings were zero and the sixth reading was 1 and was an outlier. The near lack of a blank signal does not satisfy a strict interpretation of the IUPAC method for determining the detection limits. Therefore, the standard deviation of the six peak area readings of the 0.2 ng/mL cadmium standard, 0.75 abs-sec, was used to calculate the detection limits by the IUPAC method. The slope, m, as calculated by a linear regression plot of the peak area (abs-sec) readings (shown in Table 4) of the 0.2 to 10 ng/mL cadmium standards versus their concentration is 51.5 abs-sec/(ng/mL).
(iii) If 0.75 abs-sec (sd) and 51.5 abs-sec/(ng/mL) (m) are used in Eqn. 1 ((a)(iii) of this subsection), the qualitative and quantitative detection limits as determined by the IUPAC method are:
Cld=
(3)(0.75 abs-sec)/(51.5 abs-sec/(ng/mL)= 0.044 ng/mL for the qualitative detection limit.
Cld=
(10)(0.75 abs-sec)/(51.5 abs-sec/(ng/mL)= 0.15 ng/mL for the quantitative detection limit. The qualitative and quantitative detection limits for the AAS-HGA analytical technique are 0.44 ng and 1.5 ng cadmium, respectively, for a 10 mL solution volume. These correspond, respectively, to 0.007 µg/m3 and 0.025 µg/m3 for a 60 L air volume.
(iv) The peak area (abs-sec) readings of the Cd standards from 0.2 to 40 ng/mL and the statistical analysis of the results are given in Table 4. The recommended standard working range for AAS-HGA analysis is 0.2 to 20 ng/mL. The standard of lowest concentration in the recommended working range is slightly greater than the calculated quantitative detection limit, 0.15 ng/mL. The deviation from linearity of the peak area readings of the 20 ng/mL standard, the highest concentration standard in the recommended working range, is approximately ten percent. The deviations from linearity of the peak area readings of the thirty and forty ng/mL standards are significantly greater than ten percent. As shown in Table 4, the precision of the peak area readings are satisfactory throughout the recommended working range; the relative standard deviations of the readings range from 0.025 to 0.083.
(h) Analytical method recovery for flame AAS analysis.
(i) Four sets of spiked MCEF samples were prepared by injecting 20 µL of 10, 50, 100 and 200 µg/mL dilute cadmium stock solutions on 37 mm diameter filters (part No. AAWP 037 00, Millipore Corp., Bedford, MA) with a calibrated micropipet. The dilute stock solutions were prepared by making appropriate serial dilutions of a commercially available 1,000 µg/mL cadmium standard stock solution (RICCA Chemical Co., Lot # A102) with the diluting solution (4% HNO3, 0.4% HCl). Each set contained six samples and a sample blank. The amount of cadmium in the prepared sets were equivalent to 0.1, 0.5, 1.0 and 2.0 times the TWA PEL target concentration of 5 µg/m3 for a 400 L air volume.
(ii) The air-dried spiked filters were digested and analyzed for their cadmium content by flame atomic absorption spectroscopy (AAS) following the procedure described in subsection (3) of this section. The 0.02 to 2.0 µg/mL cadmium standards (the suggested working range) were used in the analysis of the spiked filters.
(iii) The results of the analysis are given in Table 5. One result at 0.5 times the TWA PEL target concentration was an outlier and was excluded from statistical analysis. Experimental justification for rejecting it is that the outlier value was probably due to a spiking error. The coefficients of variation for the three test levels at 0.5 to 2.0 times the TWA PEL target concentration passed the Bartlett's test and were pooled.
(iv) The average recovery of the six spiked filter samples at 0.1 times the TWA PEL target concentration was 118.2% with a coefficient of variation (CV1) of 0.128. The average recovery of the spiked filter samples in the range of 0.5 to 2.0 times the TWA target concentration was 104.0% with a pooled coefficient of variation (CV1) of 0.010. Consequently, the analytical bias found in these spiked sample results over the tested concentration range was +4.0% and the OAE was ±6.0%.
(i) Analytical method recovery for AAS-HGA analysis.
(i) Three sets of spiked MCEF samples were prepared by injecting 15 µL of 5, 10 and 20 µg/mL dilute cadmium stock solutions on 37 mm diameter filters (part no. AAWP 037 00, Millipore Corp., Bedford, MA) with a calibrated micropipet. The dilute stock solutions were prepared by making appropriate serial dilutions of a commercially available certified 1,000 µg/mL cadmium standard stock solution (Fisher Chemical Co., Lot# 913438-24) with the diluting solution (4% HNO3, 0.4% HCl). Each set contained six samples and a sample blank. The amount of cadmium in the prepared sets were equivalent to 0.5, 1 and 2 times the action level TWA target concentration of 2.5 µg/m3 for a 60 L air volume.
(ii) The air-dried spiked filters were digested and analyzed for their cadmium content by flameless atomic absorption spectroscopy using a heated graphite furnace atomizer following the procedure described in subsection (3) of this section. A five-fold dilution of the spiked filter samples at 2 times the action level TWA was made prior to their analysis. The 0.05 to 20 ng/mL cadmium standards were used in the analysis of the spiked filters.
(iii) The results of the analysis are given in Table 6. There were no outliers. The coefficients of variation for the three test levels at 0.5 to 2.0 times the action level TWA PEL passed the Bartlett's test and were pooled. The average recovery of the spiked filter samples was 94.2% with a pooled coefficient of variation (CV1) of 0.043. Consequently, the analytical bias was -5.8% and the OAE was ±14.2%.
(j) Conclusions.
The experiments performed in this evaluation show the two atomic absorption analytical techniques included in this method to be precise and accurate and have sufficient sensitivity to measure airborne cadmium over a broad range of exposure levels and sampling periods.
(5) References.
(a) Slavin, W. Graphite Furnace AAS—A Source Book; Perkin-Elmer Corp., Spectroscopy Div.: Ridgefield, CT, 1984; p. 18 and pp. 83-90.
(b) Grosser, Z., Ed.; Techniques in Graphite Furnace Atomic Absorption Spectrophotometry; Perkin-Elmer Corp., Spectroscopy Div.: Ridgefield, CT, 1985.
(c) Occupational Safety and Health Administration Salt Lake Technical Center: Metal and Metalloid Particulate in Workplace Atmospheres (Atomic Absorption) (USDOL/OSHA Method No. ID-121). In OSHA Analytical Methods Manual 2nd ed. Cincinnati, OH: American Conference of Governmental Industrial Hygienists, 1991.
(d) Occupational Safety and Health Administration Salt Lake Technical Center: Metal and Metalloid Particulate in Workplace Atmospheres (ICP) (USDOL/OSHA Method No. ID-125G). In OSHA Analytical Methods Manual 2nd ed. Cincinnati, OH: American Conference of Governmental Industrial Hygienists, 1991.
(e) Windholz, M., Ed.; The Merck Index, 10th ed.; Merck & Co.: Rahway, NJ, 1983.
(f) Analytical Methods for Atomic Absorption Spectrophotometry, The Perkin-Elmer Corporation: Norwalk, CT, 1982.
(g) Slavin, W., D.C. Manning, G. Carnrick, and E. Pruszkowska: Properties of the Cadmium Determination with the Platform Furnace and Zeeman Background Correction. Spectrochim. Acta 38B:1157-1170 (1983).
(h) Occupational Safety and Health Administration Salt Lake Technical Center: Standard Operating Procedure for Atomic Absorption. Salt Lake City, UT: USDOL/OSHA-SLTC, In progress.
(i) Occupational Safety and Health Administration Salt Lake Technical Center: AAS-HGA Standard Operating Procedure. Salt Lake City, UT: USDOL/OSHA- SLTC, In progress.
(j) Mandel, J.: Accuracy and Precision, Evaluation and Interpretation of Analytical Results, The Treatment of Outliers. In Treatise On Analytical Chemistry, 2nd ed., Vol.1, edited by I. M. Kolthoff and P. J. Elving. New York: John Wiley and Sons, 1978. pp. 282-285.
(k) National Institute for Occupational Safety and Health: Documentation of the NIOSH Validation Tests by D. Taylor, R. Kupel, and J. Bryant (DHEW/NIOSH Pub. No. 77-185). Cincinnati, OH: National Institute for Occupational Safety and Health, 1977.
(l) Occupational Safety and Health Administration Analytical Laboratory: Precision and Accuracy Data Protocol for Laboratory Validations. In OSHA Analytical Methods Manual 1st ed. Cincinnati, OH: American Conference of Governmental Industrial Hygienists (Pub. No. ISBN: 0-936712-66-X), 1985.
(m) Long, G.L. and J.D. Winefordner: Limit of Detection—A Closer Look at the IUPAC Definition. Anal.Chem. 55:712A-724A (1983).
(n) American Conference of Governmental Industrial Hygienists: Documentation of Threshold Limit Values and Biological Exposure Indices. 5th ed. Cincinnati, OH: American Conference of Governmental Industrial Hygienists, 1986.
Table 1—Cd Detection Limit Study
[Flame AAS Analysis]
STD (µg/mL)
Absorbance
reading at
228.8 nm
Statistical
analysis
Reagent blank
5
2
n=6.
 
4
3
mean=3.50.
 
4
3
std dev=1.05.
 
 
 
CV=0.30.
0.001
6
6
n=6.
 
2
4
mean=5.00.
 
6
6
std dev=1.67.
 
 
 
CV=0.335.
0.002
5
7
n=6.
 
7
3
mean=5.50.
 
7
4
std dev=1.76.
 
 
 
CV=0.320.
0.005
7
7
n=6.
 
8
8
mean=7.33.
 
8
6
std dev=0.817.
 
 
 
CV=0.111.
0.010
10
9
n=6.
 
10
13
mean=10.3.
 
10
10
std dev=1.37.
 
 
 
CV=0.133.
0.020
20
23
n=6.
 
20
22
mean=20.8.
 
20
20
std dev=1.33.
 
 
 
CV=0.064.
0.050
42
42
n=6.
 
42
42
mean=42.5.
 
42
45
std dev=1.22.
 
 
 
CV=0.029.
0.10
 
84
n=3.
 
 
80
mean=82.3.
 
 
83
std dev=2.08.
 
 
 
CV=0.025.
Table 2—Cd Standard Working Range Study
[Flame AAS Analysis]
STD(µg/mL)
Absorbance
reading at
228.8 nm
Statistical
analysis
Reagent blank
5
2
n=6.
 
4
3
mean=3.50.
 
4
3
std dev=1.05.
 
 
 
CV=0.30.
0.020
20
23
n=6.
 
20
22
mean=20.8.
 
20
20
std dev=1.33.
0.050
42
42
n=6.
 
42
42
mean=42.5.
 
42
45
std dev=1.22.
 
 
 
CV=0.029.
0.10
 
84
n=3.
 
 
80
mean=82.3.
 
 
83
std dev=2.08.
 
 
 
CV=0.025.
0.20
 
161
n=3.
 
 
161
mean=160.0.
 
 
158
std dev=1.73.
 
 
 
CV=0.011.
0.50
 
391
n=3.
 
 
389
mean=391.0.
 
 
393
std dev=2.00.
 
 
 
CV=0.005.
1.00
 
760
n=3.
 
 
748
mean=753.3.
 
 
752
std dev=6.11.
 
 
 
CV=0.008.
2.00
 
1416
n=3.
 
 
1426
mean=1414.3.
 
 
1401
std dev=12.6.
 
 
 
CV=0.009.
Table 3—Cd Detection Limit Study
[AAS-HGA Analysis]
STD (ng/mL)
Peak area
readings
x 103 at
228.8 nm
Statistical analysis
Reagent blank
0
0
n=6.
 
0
1
mean=0.167.
 
0
0
std dev=0.41.
 
 
 
CV=2.45.
0.1
8
6
n=6.
 
5
7
mean=7.7.
 
13
7
std dev=2.8.
 
 
 
CV=0.366.
0.2
11
13
n=6.
 
11
12
mean=11.8.
 
12
12
std dev=0.75.
 
 
 
CV=0.064.
0.5
28
33
n=6.
 
26
28
mean=28.8.
 
28
30
std dev=2.4.
 
 
 
CV=0.083.
1.0
52
55
n=6.
 
56
58
mean=54.8.
 
54
54
std dev=2.0.
 
 
 
CV=0.037.
2.0
101
112
n=6.
 
110
110
mean=108.8.
 
110
110
std dev=3.9.
 
 
 
CV=0.036.
Table 4—Cd Standard Working Range Study
[AAS-HGA Analysis]
STD (ng/mL)
Peak area
readings
x 103 at
228.8 nm
Statistical
analysis
0.2
11
13
n=6.
 
11
12
mean=11.8.
 
12
12
std dev=0.75.
 
 
 
CV=0.064.
0.5
28
33
n=6.
 
26
28
mean=28.8.
 
28
30
std dev=2.4.
 
 
 
CV=0.083.
1.0
52
55
n=6.
 
56
58
mean=54.8.
 
54
54
std dev=2.0.
 
 
 
CV=0.037.
2.0
101
112
n=6.
 
110
110
mean=108.8.
 
110
110
std dev=3.9.
 
 
 
CV=0.036.
5.0
247
265
n=6.
 
268
275
mean=265.5.
 
259
279
std dev=11.5.
 
 
 
CV=0.044.
10.0
495
520
n=6.
 
523
513
mean=516.7.
 
516
533
std dev=12.7.
 
 
 
CV=0.025.
20.0
950
953
n=6.
 
951
958
mean=941.8.
 
949
890
std dev=25.6.
 
 
 
CV=0.027.
30.0
1269
1291
n=6.
 
1303
1307
mean=1293.
 
1295
1290
std dev=13.3.
 
 
 
CV=0.010.
40.0
1505
1567
n=6.
 
1535
1567
mean=1552.
 
1566
1572
std dev=26.6.
 
 
 
CV=0.017.
Table 5—Analytical Method Recovery
[Flame AAS Analysis]
Test level
0.5x
1.0x
 
 
2.0x
 
µg
taken
µg
found
Percent rec.
µg
taken
µg
found
Percent
rec.
µg
taken
µg
found
Percent rec.
1.00
1.0715
107.2
2.00
2.0688
103.4
4.00
4.1504
103.8
1.00
1.0842
108.4
2.00
2.0174
100.9
4.00
4.1108
102.8
1.00
1.0842
108.4
2.00
2.0431
102.2
4.00
4.0581
101.5
1.00
*1.0081
*100.8
2.00
2.0431
102.2
4.00
4.0844
102.1
1.00
1.0715
107.2
2.00
2.0174
100.9
4.00
4.1504
103.8
1.00
1.0842
108.4
2.00
2.0045
100.2
4.00
4.1899
104.7
 
n=
5
6
6
 
mean=
107.9
101.6
103.1
std dev=
0.657
1.174
1.199
CV1=
0.006
0.011
0.012
CV1
 
 
 
(pooled)=
0.010
 
 
*Rejected as an outlier - This value did not pass the outlier T-test at the 99% confidence level.
Test level 0.1x
taken
µg
found
µg
rec.
Percent
0.200
0.2509
125.5
 
0.200
0.2509
125.5
 
0.200
0.2761
138.1
 
0.200
0.2258
112.9
 
0.200
0.2258
112.9
 
0.200
0.1881
94.1
 
 
n=
6
 
 
mean=
118.2
 
 
std dev=
15.1
 
 
CV1=
0.128
 
Table 6—Analytical Method Recovery
[AAS-HGA analysis]
Test level
0.5x
1.0x
 
2.0x
ng taken
ng
found
Percent rec.
ng taken
ng
found
Percent rec.
ng
taken
ng
found
Percent rec.
75
71.23
95.0
150
138.00
92.0
300
258.43
86.1
75
71.47
95.3
150
138.29
92.2
300
258.46
86.2
75
70.02
93.4
150
136.30
90.9
300
280.55
93.5
75
77.34
103.1
150
146.62
97.7
300
288.34
96.1
75
78.32
104.4
150
145.17
96.8
300
261.74
87.2
75
71.96
95.9
150
144.88
96.6
300
277.22
92.4
 
n=
6
6
6
 
 
 
mean=
97.9
94.4
90.3
 
 
 
std dev=
4.66
2.98
4.30
 
 
 
CV1=
0.048
0.032
0.048
 
 
 
CV1
 
 
 
 
 
 
(pooled)=
0.043
 
 
 
 
(6) Instrumental Parameters for Flame AAS Analysis
Atomic Absorption Spectrophotometer
(Perkin-Elmer Model 603)
Flame: Air/Acetylene—lean, blue
Oxidant Flow: 55
Fuel Flow: 32
Wavelength: 228.8 nm
Slit: 4 (0.7 nm)
Range: UV
Signal: Concentration (4 exp)
Integration Time: 3 sec
(7) Instrumental Parameters for HGA Analysis
Atomic Absorption Spectrophotometer
(Perkin-Elmer Model 5100)
Signal Type: Zeeman AA
Slitwidth: 0.7 nm
Wavelength: 228.8 nm
Measurement: Peak Area
Integration Time: 6.0 sec
BOC Time: 5 sec BOC=Background Offset
Correction. Zeeman Graphite Furnace
(Perkin-Elmer Model HGA-600)
Step
Ramp
time
(sec)
Hold
time
(sc)
Temp.
(°C)
Argon
flow
(mL/
min)
Read
(sec)
1) Predry
5
10
90
300
 
2) Dry
30
10
140
300
 
3) Char
10
20
900
300
 
4) Cool Down
1
8
30
300
 
5) Atomize
0
5
1600
0
-1
6) Burnout
1
8
2500
300
 
[Statutory Authority: Chapter 49.17 RCW. WSR 93-21-075 (Order 93-06), § 296-62-07449, filed 10/20/93, effective 12/1/93; WSR 93-07-044 (Order 93-01), § 296-62-07449, filed 3/13/93, effective 4/27/93.]
Reviser's note: The brackets and enclosed material in the text of the above section occurred in the copy filed by the agency.